首页> 美国政府科技报告 >Results of Self-Absorption Study on the Versapor(Tradename) 3000 Filters for Radioactive Particulate Air Sampling
【24h】

Results of Self-Absorption Study on the Versapor(Tradename) 3000 Filters for Radioactive Particulate Air Sampling

机译:用于放射性颗粒空气取样的Versapor(商品名)3000过滤器的自吸收研究结果

获取原文

摘要

Since the mid-1980s, Pacific Northwest National Laboratory (PNNL) has used a value of 0.85 as the correction factor for self absorption of activity for particulate radioactive air samples collected from building exhaust for environmental monitoring. This value accounts for activity that cannot be detected by direct counting of alpha and beta particles. Emissions can be degraded or blocked by filter fibers for particles buried in the filter material or by inactive dust particles collected with the radioactive particles. These filters are used for monitoring air emissions from PNNL stacks for radioactive particles. This paper describes an effort to re-evaluate self-absorption effects in particulate radioactive air sample filters (Versapor 3000, 47 mm diameter) used at PNNL. There were two methods used to characterize the samples. Sixty samples were selected from the archive for acid digestion to compare the radioactivity measured by direct gas-flow proportional counting of filters to the results obtained after acid digestion of the filter and counting again by gas-flow proportional detection. Thirty different sample filters were selected for visible light microscopy to evaluate filter loading and particulate characteristics.