首页> 美国政府科技报告 >Precision calibration of calorimeter electronics in the D0 liquid argon/uranium particle detector
【24h】

Precision calibration of calorimeter electronics in the D0 liquid argon/uranium particle detector

机译:在D0液氩/铀粒子探测器中对量热计电子器件进行精确校准

获取原文

摘要

The ability to cross calibrate thousands of channels of detector electronics is of prime importance. This paper will describe the system used to deliver and distribute a 300 nanosecond pulse across 50,000 channels of electronics with better than 0.25% difference between channels from a location more than 200 feet away. The system is used for both cross calibration and functionality checking, (i.e., missing channels). Design of a fixed width pulse generator of high stability is presented as a key ingredient in the system's overall performance. In addition, the design of a controlled impedance distribution system is discussed. 2 refs.