Implementation of Gas Sampling Chamber and Measuring Hardware for Capnograph System Considering Thermal Noise Effect and Time Response Characteristics.

机译:考虑热噪声效应和时间响应特性的二氧化碳分析仪气体采样室和测量硬件的实现。

获取原文

摘要

Most capnograph systems that can indirectly determine the partial pressure of carbon dioxide in the blood of a patient are based on NDIR(non- dispersive infrared) absorption technology. As such an NDIR gas analyzing method requires an optical absorption chamber and signal processing hardware. Accordingly, this paper designed and implemented an NDIR type CO2 gas chamber while considering the time response characteristics and lamp chopping frequency. In addition, signal processing hardware using two infrared sources was implemented to reduce the thermal background effect. The gas chamber and signal processing hardware has been tested using temperature variation and human expiration experiment. The results showed that the system could produce a stable output signal and discerning CO2 gas concentration curve similar to a typical capnogram.