首页> 美国政府科技报告 >Method to Determine Diastolic Blood Pressure Based on Pressure Pulse Propagation in the Electronic Palpation Method.
【24h】

Method to Determine Diastolic Blood Pressure Based on Pressure Pulse Propagation in the Electronic Palpation Method.

机译:电子触诊法中基于压力脉冲传播确定舒张压的方法。

获取原文

摘要

This paper evaluates the accuracy of the pulse transit time method for determining diastolic blood pressure using intra-arterial blood pressure as a reference. First, the paper describes the method. Then it discusses two sets of measurements which were carried out to determine the accuracy of the method with cardiac operated patients and healthy young volunteers. The thus obtained accuracies were +0.7 mmHg +/- 10.7 mmHg for the cardiac patients and -6.6 mmHg +/- 10.5 mmHg for the young volunteers, respectively.

著录项