首页> 美国政府科技报告 >Fluorescence-based Sensing of 2,4,6-Trinitrotoluene (TNT) Using a Multi- channeled Poly(methyl methacrylate) (PMMA) Microimmunosensor
【24h】

Fluorescence-based Sensing of 2,4,6-Trinitrotoluene (TNT) Using a Multi- channeled Poly(methyl methacrylate) (PMMA) Microimmunosensor

机译:使用多通道聚甲基丙烯酸甲酯(pmma)微量免疫传感器对2,4,6-三硝基甲苯(TNT)进行荧光检测

获取原文

摘要

Fluorescence immunoassays employing monoclonal antibodies directed against the explosive 2,4,6-trinitrotoluene (TNT) were conducted in a multi- channel microimmunosensor. The multi-channel microimmunosensor was prepared in poly (methyl methacrylate) (PMMA) via hot embossing from a brass molding tool. The multi-channeled microfluidic device was sol-gel coated to generate a siloxane surface that provided a scaffold for antibody immobilization. AlexaFluor-cadaverine- trinitrobenzene (AlexaFluor- Cad-TNB) was used as the reporter molecule in a displacement immunoassay. The limit of detection was 1- 10 ng/mL (ppb) with a linear dynamic range that covered three orders of magnitude. In addition, antibody crossreactivity was investigated using hexahydro-1,3,5- triazine (RDX), HMX, 2,4-dinitrotoluene (DNT), 4-nitrotoluene (4-NT) and 2-amino-4,6- DNT.