Development of Biological Research Apparatus for Use in Acceleration and Deceleration Studies. Phase I. The Evaluation of Pressure Transducer Systems.

机译:开发用于加速和减速研究的生物研究仪器。第一阶段压力传感器系统的评估。

获取原文

摘要

Response measurements were made of various pressure transducers, transducer fittings, and coupling systems using pressure-step function, sine- wave pressure, and absolute calibration techniques. The accurate standard methods of transducer-system evaluation demonstrated that these systems may be treated as lumped systems. A capacitance-type transducer with a long polyvinyl catheter was used as the test system for consideration of a transducer system with more than 1 degree of freedom. Measurement of physical constants of this system were carried out and the electrical-transmission-line theory was applied to the catheter. Theoretical consideration was given to the effect of varying physical constants upon the response of the system. A damping segment, based on the theoretical concepts, improved the amplitude vs frequency response of the system and reduced acceleration artifacts.