首页> 美国政府科技报告 >Research and Development of an Incremental Microflowmeter for Biological Liquids
【24h】

Research and Development of an Incremental Microflowmeter for Biological Liquids

机译:生物液体增量微流量计的研究与开发

获取原文

摘要

This report summarizes work done in the development and evaluation of a recording incremental microflowmeter which will be applicable in studies of renal physiology, salivary gland physiology and rheologic study of lymph. The laboratory model has a stroke volume of 10 microliters and a maximum pumping rate of about 5700 microliters per minute with a demonstrated accuracy of 98 percent. Accumulated data are displayed on paper tape; both time and volume are recorded. Printout of accumulated volume occurs at 6-second intervals, and clearing occurs each minute. Elapsed time is recorded cumulatively on a minute basis. Pumping occurs when liquid level in the reservoir exceeds a critical level. Pumping rate is electronically controlled. The system requires about 175 watts of electrical power at 120 volts and 60 hertz. The laboratory model incremental microflowmeter was mounted in a 11 x 13 x 22-inch cabinet and weighs about 35 pounds. (Author)

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号