High-Temperature Mass Spectrometry. Volume II. Knudsen Cell Assembly for Herzog-Mattauch Type Mass Spectrometer,

机译:高温质谱。第二卷。用于Herzog-mattauch型质谱仪的Knudsen电池组件,

获取原文

摘要

A Knudsen cell assembly that is capable of operating up to 2750K for use with the high-resolution Herzog-Mattauch type double-focusing mass spectrometer is described. The design of this Knudsen cell assembly is unusual,because the effusin orifice is situated on the curved side of the cylindrical cell and produces a molecular beam coaxial with the ion beam. The performance of this device was checked by measuring the heats of vaporization of silver and carbon. The third-law and second-law heat of sublimation of silver was measured. Both of these heats of sublimation are in agreement with those reported in the literature. (Author)