首页> 美国政府科技报告 >Comparison of Total White Blood Cell Counts on the Technicon H6000 and Coulter Counter Model ZBI in an Occupational Health Program
【24h】

Comparison of Total White Blood Cell Counts on the Technicon H6000 and Coulter Counter Model ZBI in an Occupational Health Program

机译:Technicon H6000和Coulter计数器模型ZBI在职业健康计划中总白细胞计数的比较

获取原文

摘要

White blood cell (WBC) counts obtained using a Technicon H6000 in a large occupational health program were compared with counts obtained on identical samples using a Counter Counter Model ZBI. To assure quality control and to assess the comparability of results obtained on the two machines, a 25% systematic sample (N = 586) of analyses performed on the Technicon were independently and blindly performed on the Coulter Counter. The Pearson product-moment-correlation between WBC counts from the two machines was r = .94 (p less than or = .0001). The mean WBC count was 6,664 cells/cu.mm for the Technicon and 6,790 cells/cu.mm for the Coulter Counter. The estimated prevalence of leukopenia (WBC counts less than or = 4,500 cells/cu.mm) was 10.6% using the Technicon and 9.7% using the Coulter Counter. The results demonstrate that the two machines provide results which can be directly compared without corrections.