首页> 美国政府科技报告 >SQUID (Superconducting Quantum Interference Device) Arrays for Simultaneous Magnetic Measurements: Calibration and Source Localization Performance
【24h】

SQUID (Superconducting Quantum Interference Device) Arrays for Simultaneous Magnetic Measurements: Calibration and Source Localization Performance

机译:用于同时磁测量的sQUID(超导量子干涉装置)阵列:校准和源定位性能

获取原文

摘要

Recently developed small arrays of SQUID-based magnetic sensors can, if appropriately placed, locate the position of a confined biomagnetic source without moving the array. The authors present a technique with a relative accuracy of about 2% for calibrating such sensors having detection coils with the geometry of a second-order gradiometer. The effects of calibration error and magnetic noise on the accuracy of locating an equivalent current dipole source in the human brain are investigated for 5- and 7-sensor probes and for a pair of 7-sensor probes. With a noise level of 5% of peak signal, uncertainties of about 20% in source strength and depth for a 5-sensor probe are reduced to 8% for a pair of 7-sensor probes, and uncertainties of about 15 mm in lateral position are reduced to 1 mm, for the configuration considered.