首页> 美国政府科技报告 >Particle-Sizing System for Scanning Electron Microscope Images of Solid-Propellant Combustion Exhaust.
【24h】

Particle-Sizing System for Scanning Electron Microscope Images of Solid-Propellant Combustion Exhaust.

机译:固体推进剂燃烧废气扫描电子显微镜图像粒度测量系统。

获取原文

摘要

Accurate measurement of particle size distribution of rocket motor exhausts is essential for predicting the combustion efficiency and infrared plume signature. This thesis presents an automated method for extracting particle size distribution from scanning electron microscope (SEM) images. The SEM images were taken off a filter paper placed at the end of a collection probe inserted into the exhaust plume. The automated SEM extraction system consists of n IBM AT-based computer system fitted with a 512 x 480 pixel frame grabber. Photographic images taken off the SEM are acquired via a vidicon camera. A C language program was written to control the hardware and automate the extraction process. A threshold is first applied to the digitized image and the resulting binary image is subjected to object segmentation. Each object is then sized and the distribution from one or more images can be plotted. The main bulk of this thesis document the software specially written to undertake this set of tasks. Results obtained were compared with that from a Malvern MasterSizer particle sizer and found to be favorable. Particles as small as 1/8 microns have been successfully sized.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号