首页> 美国政府科技报告 >Non-Invasive Detection of Increased Intracranial Pressure Using AcousticalTechniques
【24h】

Non-Invasive Detection of Increased Intracranial Pressure Using AcousticalTechniques

机译:使用声学技术非侵入性检测增加的颅内压

获取原文

摘要

This study was undertaken to determine the relationships between intracranialpressure (ICP) and the acoustic response of the head. Potentially fatal increases in ICP can occur in soldiers afflicted with Acute Mountain Sickness at high-altitude, yet there is no reliable and noninvasive procedure to monitor ICP changes in uncontrolled conditions. In this investigation, an ovine animal model was used in which incremental increases in ICP were elicited and directly measured through intraventricular cannulae. At each ICP increment, a vibration source elicited a flexural response of the animal's head, and this acoustic response was measured at four locations on the skull using accelerometers. Spectral analysis of the response showed numerous changes in proportion to ICP up to roughly 20 CmH2O above normal; a clinically significant range. For example, changes in magnitude and phase at frequencies between 4 and 7 kHz each correlated well (> 0.92) with ICP across the study group. Thus, alterations in ICP give rise to measurable acoustic changes that can potentially be used to noninvasively monitor afflicted soldiers in uncontrolled environments. Altitude, Cerebral edema, Acoustics, Mountain sickness, RAD III.