A liquid chromatography-electrochemistry (LC-EC) method is described for thedetermination of basal acetylcholine (ACh) in microdialysate from the striatum of freely moving rats. This method is based on the separation of ACh and choline (Ch) by microbore liquid chromatography followed by passage of the effluent through a post-column immobilized enzyme reactor (IMER), containing acetyicholinesterase (AChE) and choline oxidase (ChO), and then the electrochemical detection of the hydrogen peroxide produced. Instead of the conventional platinum electrode generally used for the anodic detection of hydrogen peroxide, a peroxidase-redox polymer modified glassy carbon electrode operated at + 100 mV vs. Ag/AgCl has been used to detect the reduction of hydrogen peroxide. With this method, a detection limit of 10 fmol (injected) for ACh (S/N = 3:1) was obtained and the basal ACh concentration in striatal microdialysate was determined without using esterase inhibitors.