摘要

The suitability of a new technology single-monochromator diode array spectroradiometer for UV-radiation safety measurements, in particular for sunbed measurements, was evaluated. The linearity, cosine response, temperature response, wavelength scale, stray-light and slit function of the spectroradiometer were determined and their effects on the measurement accuracy evaluated. The main error sources were stray-light and nonideal cosine response, for which correction methods are presented. Without correction, the stray-light may reduce the accuracy of the measurement excessively, particularly in the UV-B range. The expanded uncertainty of the corrected UV measurements is estimated to be 14%, which is confirmed with the comparative measurements carried out with a well-characterized double-monochromator spectroradiometer. The measurement accuracy is sufficient for sunbed measurements, provided that all corrections described above have been done and the user of the instrument has a good understanding of the instrument's operating principles and potential error sources. If these requirements are met, the tested spectroradiometer improves and facilitates market surveillance field measurements of sunbeds.
机译:评估了一种新技术的单单色器二极管阵列光谱辐射仪对紫外线辐射安全性测量,特别是日光浴测量的适用性。确定了光谱仪的线性,余弦响应,温度响应,波长范围,杂散光和狭缝功能,并评估了它们对测量精度的影响。主要的误差源是杂散光和非理想的余弦响应,为此提出了校正方法。如果不进行校正,杂散光可能会严重降低测量精度,尤其是在UV-B范围内。校正后的UV测量的扩展不确定度估计为14%,这可以通过使用特征明确的双单色仪光谱仪进行的比较测量来确认。只要进行了上述所有校正,并且仪器的用户对仪器的工作原理和潜在的误差源有很好的了解,那么对于日光浴测量,测量精度就足够了。如果满足这些要求,则经过测试的光谱仪将改善并促进市场对日光浴床的现场监测。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号