首页> 外文期刊>Physiological measurement >Comment on 'Comparative analysis of the isovolume calibration method for non-invasive respiratory monitoring techniques based on area transduction versus circumference transduction using the connected cylinders model' (2011 Physiol. Meas. 32 1265-74)
【24h】

Comment on 'Comparative analysis of the isovolume calibration method for non-invasive respiratory monitoring techniques based on area transduction versus circumference transduction using the connected cylinders model' (2011 Physiol. Meas. 32 1265-74)

机译:关于“使用连接的气瓶模型的基于面积转换与周向转换的无创呼吸监测技术的等体积校准方法的比较分析”评论(2011 Physiol。Meas。32 1265-74)

获取原文
获取原文并翻译 | 示例

摘要

An analysis introduced by the authors in 2011 examining the robustness of the isovolume method for the calibration of the respiratory inductive plethysmograph based on the connected cylinders particular model of Konno and Mead's generalized two-compartment model of respiration is extended. It is demonstrated that extending this to a more physically realistic geometrical model, termed the connected prismatic elliptical segments model, does not enhance the earlier analysis, and that the analysis can easily be proven to cover all area-based transduction sensors, irrespective of the actual geometry of the compartments.
机译:作者于2011年提出的分析方法,基于等高线法,基于Konno气瓶连接模型和Mead的广义两室呼吸模型,对等体积方法进行的呼吸感应体积描记器校准的鲁棒性进行了分析。证明了将其扩展到更物理逼真的几何模型(称为连接的棱柱形椭圆段模型)并不能增强早期的分析,并且可以容易地证明该分析涵盖了所有基于面积的换能传感器,无论实际情况如何。隔室的几何形状。