Specific heat capacity of freshly excised prostate specimens

机译:新鲜切除的前列腺标本的比热容

获取原文
获取原文并翻译 | 示例
           

摘要

The specific heat capacity of tissue is a critical parameter for thermal therapies that act over a long period of time. It is also critical for thermoacoustic signal generation. We present ex vivo measurements of specific heat capacity performed by a dual-pin probe with tight temperature control of the specimen. One 30 mm × 1.28 mm probe heats steadily for 30 s, while another measurement probe measures temperatures 6 mm away from the center of the heater probe. Specific heat values ranging from 2.9 to 4 J cm ~(-3) °C ~(-1) were measured on 20 lobes from ten fresh prostate specimens with varying degrees of cancerous involvement as confirmed by histology.
机译:组织的比热容是长时间起作用的热疗法的关键参数。这对于产生热声信号也至关重要。我们介绍了通过双针探针对样品进行严格温度控制的比热容的离体测量。一个30 mm×1.28 mm的探头稳定加热30 s,而另一个测量探头则在距加热探头中心6 mm处测量温度。如组织学所证实,在十个新鲜前列腺标本的20个叶上测量了比热值,范围为2.9至4 J cm〜(-3)°C〜(-1)。