首页> 外文期刊>Physics in medicine and biology. >Contralateral breast doses measured by film dosimetry: tangential techniques and an optimized IMRT technique.
【24h】

Contralateral breast doses measured by film dosimetry: tangential techniques and an optimized IMRT technique.

机译:通过胶片剂量法测量对侧乳房的剂量:切线技术和优化的IMRT技术。

获取原文
获取原文并翻译 | 示例
           

摘要

The contralateral breast (CLB) doses for three tangential techniques were characterized by using a female thorax phantom and GafChromic EBT film. Dose calculations by the pencil beam and collapsed cone algorithms were included for comparison. The film dosimetry reveals a highly inhomogeneous dose distribution within the CLB, and skin doses due to the medial fields that are several times higher than the interior dose. These phenomena are not correctly reproduced by the calculation algorithms. All tangential techniques were found to give a mean CLB dose of approximately 0.5 Gy. All wedged fields resulted in higher CLB doses than the corresponding open fields, and the lateral open fields resulted in higher CLB doses than the medial open fields. More than a twofold increase in the mean CLB dose from the medial open field was observed for a 90 degrees change of the collimator orientation. Replacing the physical wedge with a virtual wedge reduced the mean dose to the CLB by 35% and 16% for the medial and lateral fields, respectively. Lead shielding reduced the skin dose for a tangential technique by approximately 50%, but the mean CLB dose was only reduced by approximately 11%. Finally, a technique based on open medial fields in combination with several IMRT fields is proposed as a technique for minimizing the CLB dose. With and without lead shielding, the mean CLB dose using this technique was found to be 0.20 and 0.27 Gy, respectively.
机译:通过使用女性胸模和GafChromic EBT胶片来表征三种切线技术的对侧乳房(CLB)剂量。为了进行比较,还包括了笔形光束的剂量计算和塌陷锥算法。薄膜剂量测定法显示出CLB内的剂量分布高度不均匀,并且由于内侧电场造成的皮肤剂量比内部剂量高出几倍。这些现象不能通过计算算法正确再现。发现所有切线技术的平均CLB剂量约为0.5 Gy。所有楔形区域产生的CLB剂量均高于相应的开放区域,而外侧区域产生的CLB剂量高于内侧区域。对于准直器方向90度的变化,观察到来自内侧视野的平均CLB剂量增加了两倍以上。用虚拟楔代替物理楔可使内侧和外侧场的CLB平均剂量分别减少35%和16%。铅屏蔽使切线技术的皮肤剂量减少了约50%,但平均CLB剂量仅减少了约11%。最后,提出了一种基于开放式中场并结合多个IMRT场的技术,作为最小化CLB剂量的技术。使用和不使用铅屏蔽,使用该技术的平均CLB剂量分别为0.20和0.27 Gy。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号