Modeling, validation and application of a mathematical tissue-equivalent breast phantom for linear slot-scanning digital mammography.

机译:用于线性缝隙扫描数字乳腺摄影的数学等效组织乳腺模型的建模,验证和应用。

获取原文
获取原文并翻译 | 示例
           

摘要

This paper presents a mathematical tissue-equivalent breast phantom for linear slot-scanning digital mammography. A recently developed prototype linear slot-scanning digital mammography system was used for model validation; image quality metrics such as image contrast and contrast-to-noise ratio were calculated. The results were in good agreement with values measured using a physical breast-equivalent phantom designed for mammography. The estimated pixel intensity of the mathematical phantom, the analogue-to-digital conversion gain and the detector additive noise showed good agreement with measured values with correlation of nearly 1. An application of the model, to examine the feasibility of using a monochromatic filter for dose reduction and improvement of image quality in slot-scanning digital mammography, is presented.
机译:本文提出了一种用于线性缝隙扫描数字化乳腺X线摄影的数学等效组织乳腺模型。最近开发的原型线性缝隙扫描数字乳腺X线摄影系统用于模型验证;计算了图像质量指标,例如图像对比度和对比度噪声比。结果与使用为乳房X线摄影术设计的等效于乳房的物理体模所测得的值高度吻合。数学体模的估计像素强度,模数转换增益和检测器相加噪声与相关值接近1的测量值显示出良好的一致性。该模型的应用程序,用于检验使用单色滤光片进行测量的可行性。提出了在缝隙扫描数字化乳腺X射线摄影术中减少剂量并改善图像质量。