首页> 外文期刊>Physics in medicine and biology. >Evaluation of stray radiofrequency radiation emitted by electrosurgical devices
【24h】

Evaluation of stray radiofrequency radiation emitted by electrosurgical devices

机译:评估电外科设备发出的杂散射频辐射

获取原文
获取原文并翻译 | 示例
           

摘要

Electrosurgery refers to the passage of a high-frequency, high-voltage electrical current through the body to achieve the desired surgical effects. At the same time, these procedures are accompanied by a general increase of the electromagnetic field in an operating room that may expose both patients and personnel to relatively high levels of radiofrequency radiation. In the first part of this study, we have taken into account the radiation emitted by different monopolar electrosurgical devices, evaluating the electromagnetic field strength delivered by an electrosurgical handle and straying from units and other electrosurgical accessories. As a summary, in the worst case a surgeon's hands are exposed to a continuous and pulsed RF wave whose magnetic field strength is 0.75 A m(-1) (E-field 400 V m(-1)). Occasionally stray radiation may exceed ICNIRP's occupational exposure guidelines, especially close to the patient return plate. In the second part of this paper, we have analysed areas of particular concern to prevent electromagnetic interference with some life-support devices (ventilators and electrocardiographic devices), which have failed to operate correctly. Most clinically relevant interference occurred when an electrosurgery device was used within 0.3 m of medical equipment. In the appendix, we suggest some practical recommendations intended to minimize the potential for electromagnetic hazards due to therapeutic application of RF energy.
机译:电外科手术是指高频,高压电流通过人体以实现所需的手术效果。同时,这些过程伴随着手术室中电磁场的普遍增加,这可能使患者和人员都暴露于相对较高水平的射频辐射中。在本研究的第一部分中,我们考虑了不同单极电外科手术设备发出的辐射,评估了电外科手术手柄传递的电磁场强度以及从设备和其他电外科手术附件中散布的电磁场强度。总之,在最坏的情况下,外科医生的手会暴露在磁场强度为0.75 A m(-1)(电场为400 V m(-1))的连续脉冲RF波中。有时,杂散辐射可能会超出ICNIRP的职业暴露准则,尤其是靠近患者回流板的地方。在本文的第二部分中,我们分析了需要特别关注的领域,以防止对某些无法正常运行的生命支持设备(呼吸机和心电图设备)产生电磁干扰。当在0.3 m的医疗设备中使用电外科手术设备时,发生了大多数与临床相关的干扰。在附录中,我们提出了一些实用的建议,旨在最大程度地减少由于射频能量的治疗应用而引起的电磁危害的可能性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号