Respiratory gating based on external surrogates is performed in many clinics. We have developed a new technique for treatment verification using an electronic portal imaging device (EPID) in cine mode for gated 3D conformal therapy. Implanted radiopaque fiducial markers inside or near the target are required for this technique. The markers are contoured on the planning CT set, enabling us to create digitally reconstructed radiographs (DRRs) for each treatment beam. During the treatment, a sequence of EPID images can be acquired without disrupting the treatment. Implanted markers are visualized in the images and their positions in the beam's eye view are calculated off-line and compared to the reference position by matching the field apertures in corresponding EPID and DRR images. The precision of the patient setup, the placement of the beam-gating window, as well as the residual tumour motion can be assessed for each treatment fraction. This technique has been demonstrated with a case study patient, who had three markers implanted in his liver. For this patient, the intra-fractional variation of all marker positions in the gating window had a 95% range of 4.8 mm in the SI direction (the primary axis of motion). This was about the same (5 mm) as the residual motion considered in the planning process. The inter-fractional variation of the daily mean positions of the markers, which indicates the uncertainty in the set-up procedure, was within +8.3 mm/-4.5 mm (95% range) in the SI direction for this case.
机译:许多诊所都采用基于外部替代物的呼吸门控。我们开发了一种新技术,用于以电影模式使用电子门禁成像设备(EPID)对门控3D保形疗法进行治疗验证。该技术需要在靶标内或靶标附近植入不透射线的基准标记。标记在计划的CT集合上绘制轮廓,使我们能够为每个治疗束创建数字重建的X射线照片(DRR)。在治疗期间,可以在不中断治疗的情况下获取一系列EPID图像。在图像中可视化了已植入的标记,并离线计算了它们在光束眼睛视图中的位置,并通过匹配相应EPID和DRR图像中的视场光圈,将其与参考位置进行了比较。可以针对每个治疗分数评估患者设置的精度,射束选通窗口的放置以及残留的肿瘤运动。案例研究的病人已经证明了这种技术,他的肝脏中植入了三个标记物。对于此患者,门控窗口中所有标记位置的分数内变化在SI方向(运动的主轴)上的95%范围为4.8 mm。这与计划过程中考虑的残余运动大约相同(5毫米)。在这种情况下,标记的日平均位置的分数间变化(指示设置过程中的不确定性)在SI方向上在+8.3 mm / -4.5 mm(95%范围)内。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号