Post-irradiation colouration of Gafchromic EBT radiochromic film.

机译:Gafchromic EBT放射变色膜的辐照后着色。

获取原文
获取原文并翻译 | 示例
       

摘要

Gafchromic EBT (International Specialty Products, NJ, USA), radiochromic film is one of the newest radiation-induced auto-developing x-ray analysis films available for therapeutic radiation dosimetry in radiotherapy applications. Part of any radiochromic film product which undergoes a polymerization reaction for automatic darkening is an associated post-irradiation colouration whereby the film continues to darken after irradiation has ceased. The Gafchromic EBT film has been shown to produce an approximate 6% to 9% increase in post-irradiation optical density within the first 12 h of irradiation within the 1 Gy to 5 Gy dose range. This is compared to approximately 13%, 15% and 19% for MD-55-2, XR type T and HS radiochromic film, respectively. It is also shown that the EBT film's post-irradiation growth stabilizes to within 1% within the first 6 h. Thus EBT provides a reduced post-irradiation growth effect. However, to increase the accuracy of the film analysis, it is recommended that films be left for a significant period (at least 6 h) before the analysis is performed to provide a high level of accuracy. Also, calibration films must be read out with the same post-irradiation time to further enhance the accuracy of dosimetry.
机译:Gafchromic EBT(国际特种产品,美国新泽西州),放射致变色膜是用于放射治疗应用中放射治疗剂量测定的最新辐射诱导自动显影X射线分析膜之一。进行聚合反应以自动变黑的任何辐射致变色膜产品的一部分是与之相关的辐照后着色,由此膜在辐照停止后会继续变黑。 Gafchromic EBT膜已显示在1 Gy至5 Gy剂量范围内的照射的前12小时内,照射后的光密度增加约6%至9%。相比之下,MD-55-2,XR T型和HS放射致变色膜分别约为13%,15%和19%。还显示,在最初的6小时内,EBT膜的辐照后生长稳定在1%以内。因此,EBT提供了降低的辐照后生长效果。但是,为了提高胶片分析的准确性,建议在执行分析之前将胶片放置相当长的一段时间(至少6小时),以提供较高的准确性。同样,必须在相同的辐照后时间读出校准膜,以进一步提高剂量测定的准确性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号