Characterization of a 2D ion chamber array for the verification of radiotherapy treatments

机译:二维离子室阵列的表征,用于验证放射治疗

获取原文
获取原文并翻译 | 示例
       

摘要

In this study we investigated the characteristics of a commercial ion chamber array and its performance in the verification of radiotherapy plans. The device was the 2D Array Seven29(TM) model (PTW, Freiburg, Germany). This is a two-dimensional detector array with 729 ionization chambers uniformly arranged in a 27 x 27 matrix with an active area of 27 x 27 cm(2). The detector short-, medium- and long-term reproducibility have been tested through an extensive set of repeated measurements. Short-term reproducibility was well within 0.2%. Medium- and long-term reproducibility were within 1%, including setup reproducibility errors and linac output fluctuations. Dose linearity was also assessed. The system response to dose was verified to be linear within the range 2-500 MU. Output factors matched very well pinpoint chamber measurements performed in the same experimental conditions with a maximum local percentage difference of 0.4%. Furthermore, the 2D Array sensitivity to millimetric collimator positional changes and to perturbation effect of irradiated area was tested. The comparison with ion chamber data carried out in water was very satisfying. Finally, measurements of wedge-modulated fields and IMRT beam sequence matched very well ion chamber dose profiles acquired in a water tank. The extensive tests performed in this investigation show that the 2D Array Seven29 is a reliable and accurate dosimeter and that it could be a useful toot for the quality assurance and the verification of radiotherapy plans.
机译:在这项研究中,我们调查了商业离子室阵列的特性及其在放射治疗计划验证中的性能。该设备是2D Array Seven29™型号(PTW,德国弗赖堡)。这是一个二维检测器阵列,具有729个电离室,均匀分布在27 x 27矩阵中,有效面积为27 x 27 cm(2)。该检测器的短期,中期和长期可重复性已经通过大量重复测量进行了测试。短期再现性在0.2%以内。中长期重复性均在1%以内,包括设置重复性误差和直线加速器输出波动。还评估了剂量线性。验证了系统对剂量的响应在2-500 MU范围内是线性的。输出因子与在相同实验条件下进行的精确腔室测量非常匹配,最大局部百分比差为0.4%。此外,测试了二维阵列对毫米准直仪位置变化和对辐照区微扰效应的敏感性。与水中离子室数据的比较非常令人满意。最后,楔形调制场和IMRT射束序列的测量值与在水箱中获得的非常好的离子室剂量分布图相匹配。在这项研究中进行的大量测试表明,二维阵列Seven29是可靠且准确的剂量计,对于质量保证和放射治疗计划的验证,它可能是有用的嘟嘟声。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号