摘要

Solid state pellets (1 mm thick) for electron spin resonance (ESR) dosimetry were made using ammonium tartrate as the radiation-sensitive substance. Their behaviour was experimentally investigated as a function of dose with 60Co gamma rays. The calibration function obtained permits measurements of absorbed dose in the 2-50 Gy range, with a combined uncertainty of +/-4%. The lowest detectable dose was about 0.5 Gy. These properties are comparable with or even better than those of ESR dosimeters made from other materials. The time stability of the ESR signal of ammonium tartrate dosimeters at different storage conditions after irradiation was studied. A rather complex behaviour was observed, which suggests that more species of free radicals are produced by radiation and that migration processes may be effective. No dependence of the response on beam quality was found for high-energy photon and electron beams produced by a linear accelerator used in radiotherapy, whereas dose was underestimated with low-energy x-rays.
机译:使用酒石酸铵作为辐射敏感物质,制备了用于电子自旋共振(ESR)剂量测定的固态颗粒(1毫米厚)。实验研究了它们的行为与60Coγ射线剂量的关系。所获得的校准功能可以测量2-50 Gy范围内的吸收剂量,总不确定度为+/- 4%。最低可检测剂量为约0.5 Gy。这些性能与其他材料制成的ESR剂量计相当甚至更好。研究了酒石酸铵剂量计在不同储存条件下的ESR信号的时间稳定性。观察到相当复杂的行为,这表明通过辐射产生了更多种类的自由基,并且迁移过程可能是有效的。对于放疗中使用的线性加速器产生的高能光子和电子束,未发现响应对束质量的依赖性,而低能X射线低估了剂量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号