Brachytherapy is a very successful treatment option for many different tumors. Consequently the quality assurance ( QA) of afterloaders is of special interest. For many quality checks verification films are essential. Because currently in many clinics the film development units are replaced by computed radiography systems, the question arises of how to perform QA for afterloaders without conventional radiography films. In this note the use of a new designed multislit phantom demonstrates the possibility of performing QA checks with high sensitivity computed radiography systems. Because the presented phantom allows a method for imaging the radiation source it might be useful in the development of further QA techniques.
机译:近距离放射疗法是许多不同肿瘤的非常成功的治疗选择。因此,后装载机的质量保证(QA)引起了特别关注。对于许多质量检查而言,验证胶片至关重要。因为当前在许多诊所中,胶片显影单元已被计算机射线照相系统所取代,所以出现了一个问题,即如何在没有常规射线照相胶片的情况下对后装机进行质量检查。在本说明中,使用新设计的多缝幻像仪演示了使用高灵敏度计算机放射成像系统执行QA检查的可能性。由于所提供的体模允许使用一种对辐射源成像的方法,因此在开发进一步的QA技术时可能会很有用。