Radiation risk assessment in neonatal radiographic examinations of the chest and abdomen: a clinical and Monte Carlo dosimetry study

机译:新生儿胸部和腹部X光检查中的放射风险评估:临床和蒙特卡洛剂量学研究

获取原文
获取原文并翻译 | 示例
           

摘要

Seeking to assess the radiation risk associated with radiological examinations in neonatal intensive care units, thermo-luminescence dosimetry was used for the measurement of entrance surface dose (ESD) in 44 AP chest and 28 AP combined chest-abdominal exposures of a sample of 60 neonates. The mean values of ESD were found to be equal to 44 +/- 16 mu Gy and 43 +/- 19 mu Gy, respectively. The MCNP-4C2 code with a mathematical phantom simulating a neonate and appropriate x-ray energy spectra were employed for the simulation of the AP chest and AP combined chest-abdominal exposures. Equivalent organ dose per unit ESD and energy imparted per unit ESD calculations are presented in tabular form. Combined with ESD measurements, these calculations yield an effective dose of 10.2 +/- 3.7 mu Sv, regardless of sex, and an imparted energy of 18.5 +/- 6.7 mu J for the chest radiograph. The corresponding results for the combined chest-abdominal examination are 14.7 +/- 7.6 mu Sv (males)/17.2 +/- 7.6 mu Sv ( females) and 29.7 +/- 13.2 mu J. The calculated total risk per radiograph was low, ranging between 1.7 and 2.9 per million neonates, per film, and being slightly higher for females. Results of this study are in good agreement with previous studies, especially in view of the diversity met in the calculation methods.
机译:为了评估新生儿重症监护病房与放射学检查相关的放射风险,采用热发光剂量法测量了60例新生儿的44个AP胸部和28个AP合并胸腹暴露的入射表面剂量(ESD) 。发现ESD的平均值分别等于44 +/- 16μGy和43 +/- 19μGy。使用带有模拟新生儿的数学模型的MCNP-4C2代码和适当的X射线能谱来模拟AP胸部和AP组合的腹部与腹部的暴露。单位ESD的等效器官剂量和单位ESD计算所赋予的能量以表格形式显示。结合ESD测量,这些计算得出的有效剂量为10.2 +/- 3.7μSv,与性别无关,而胸部X光片的能量为18.5 +/- 6.7μJ。胸部-腹部综合检查的相应结果为:男性14.7 +/- 7.6μSv(女性)/女性17.2 +/- 7.6 mu Sv和29.7 +/- 13.2μJ。每部电影,每百万名新生儿介于1.7至2.9之间,而女性则略高。这项研究的结果与以前的研究非常吻合,特别是考虑到计算方法的多样性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号