首页> 外文期刊>Physics in medicine and biology. >A modified method of calculating the lateral build-up ratio for small electron fields
【24h】

A modified method of calculating the lateral build-up ratio for small electron fields

机译:一种计算小电子场横向堆积比的改进方法

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

This note outlines an improved method of calculating dose per monitor unit values for small electron fields using Khan's lateral build-up ratio (LBR). This modified method obtains the LBR directly from the ratio of measured, surface normalized, electron beam percentage depth dose curves. The LBR calculated using this modified method more accurately accounts for the change in lateral scatter with decreasing field size. The LBR is used along with Khan's dose per monitor unit formula to calculate dose per monitor unit values for a set of small fields. These calculated dose per monitor unit values are compared to measured values to within 3.5% for all circular fields and electron energies examined. The modified method was further tested using a small triangular field. A maximum difference of 4.8% was found.