The positron emission mammography/tomography breast imaging and biopsy system (PEM/PET): design, construction and phantom-based measurements

机译:正电子发射乳腺摄影/断层摄影术乳房成像和活检系统(PEM / PET):设计,构造和基于幻像的测量

获取原文
获取原文并翻译 | 示例

摘要

Tomographic breast imaging techniques can potentially improve detection and diagnosis of cancer in women with radiodense and/or fibrocystic breasts. We have developed a high-resolution positron emission mammography/tomography imaging and biopsy device (called PEM/PET) to detect and guide the biopsy of suspicious breast lesions. PET images are acquired to detect suspicious focal uptake of the radiotracer and guide biopsy of the area. Limited-angle PEM images could then be used to verify the biopsy needle position prior to tissue sampling. The PEM/PET scanner consists of two sets of rotating planar detector heads. Each detector consists of a 4 x 3 array of Hamamatsu H8500 flat panel position sensitive photomultipliers (PSPMTs) coupled to a 96 x 72 array of 2 x 2 x 15 mm(3) LYSO detector elements (pitch = 2.1 mm). Image reconstruction is performed with a three-dimensional, ordered set expectation maximization (OSEM) algorithm parallelized to run on a multiprocessor computer system. The reconstructed field of view (FOV) is 15 x 15 x 15 cm(3). Initial phantom-based testing of the device is focusing upon its PET imaging capabilities. Specifically, spatial resolution and detection sensitivity were assessed. The results from these measurements yielded a spatial resolution at the center of the FOV of 2.01 +/- 0.09 mm (radial), 2.04 +/- 0.08 mm (tangential) and 1.84 +/- 0.07 mm (axial). At a radius of 7 cm from the center of the scanner, the results were 2.11 +/- 0.08 mm (radial), 2.16 +/- 0.07 mm (tangential) and 1.87 +/- 0.08 mm (axial). Maximum system detection sensitivity of the scanner is 488.9 kcps mu Ci(-1) ml(-1) (6.88%). These promising findings indicate that PEM/PET may be an effective system for the detection and diagnosis of breast cancer.
机译:乳腺断层扫描成像技术可以潜在地改善具有放射致密性和/或纤维囊性乳房的女性的癌症检测和诊断。我们已经开发出高分辨率的正电子发射乳房X线摄影/断层摄影成像和活检设备(称为PEM / PET),以检测和指导可疑乳腺病变的活检。采集PET图像以检测放射性示踪剂的可疑局灶性摄取并指导该区域的活检。然后,可以使用有限角度的PEM图像在组织采样之前验证活检针的位置。 PEM / PET扫描仪由两组旋转的平面探测器头组成。每个检测器均由4 x 3的Hamamatsu H8500平板位置敏感光电倍增器(PSPMT)阵列组成,该阵列与2 x 2 x 15毫米(3)LYSO检测器元件(间距= 2.1毫米)的96 x 72阵列相连。图像重建是通过并行执行在多处理器计算机系统上运行的三维有序集合期望最大化(OSEM)算法执行的。重建视场(FOV)为15 x 15 x 15 cm(3)。最初基于幻像的设备测试集中在其PET成像功能上。具体而言,评估了空间分辨率和检测灵敏度。这些测量的结果在FOV的中心产生了2.01 +/- 0.09毫米(径向),2.04 +/- 0.08毫米(切向)和1.84 +/- 0.07毫米(轴向)的空间分辨率。在距扫描仪中心7厘米的半径处,结果为2.11 +/- 0.08毫米(径向),2.16 +/- 0.07毫米(切向)和1.87 +/- 0.08毫米(轴向)。扫描仪的最大系统检测灵敏度为488.9 kcps mu Ci(-1)ml(-1)(6.88%)。这些有希望的发现表明,PEM / PET可能是检测和诊断乳腺癌的有效系统。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号