首页> 外文期刊>Physics in medicine and biology. >Study and evaluation of the Siemens virtual wedge factor: dosimetric monitor system and variable field effects
【24h】

Study and evaluation of the Siemens virtual wedge factor: dosimetric monitor system and variable field effects

机译:西门子虚拟楔形因子的研究和评估:剂量监测系统和可变场效应

获取原文
获取原文并翻译 | 示例
           

摘要

In the year 1997 Siemens introduced the virtual wedge in its accelerators. The idea was that a dose profile similar to that of a physical wedge can be obtained by moving one of the accelerator jaws at a constant speed while the dose rate is changing. This work explores the observed behaviour of virtual wedge factors. A model is suggested which takes into account that at any point in time, when the jaw moves, the dose at a point of interest in the phantom is not only due to the direct beam. It also depends on the scattered radiation in the phantom, the head scatter and the behaviour of the monitoring system of the accelerator. Measurements are performed in a Siemens Primus accelerator and compared to the model predictions. It is shown that the model agrees reasonably well with measurements spanning a wide range of conditions. A strong dependence of virtual wedge factors on the dosimetric board has been confirmed and an explanation has been given on how the balance between different contributions is responsible for virtual wedge factors values.
机译:在1997年,西门子在其加速器中引入了虚拟楔子。这个想法是,当剂量率改变时,通过以恒定的速度移动一个加速器钳口,可以获得类似于物理楔形的剂量分布。这项工作探索了虚拟楔形因子的观察到的行为。建议使用一种模型,该模型考虑到在下颚移动的任何时间点,体模中感兴趣点的剂量不仅是由于直接射束引起的。它还取决于体模中的散射辐射,头部散射以及加速器监控系统的行为。测量在Siemens Primus加速器中进行,并与模型预测值进行比较。结果表明,该模型与广泛条件下的测量结果相当吻合。确认了虚拟楔形因子对剂量测定板的强烈依赖性,并给出了关于不同贡献之间的平衡如何导致虚拟楔形因子值的解释。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号