Analysis of B1 field profiles and SAR values for multi-strut transverse electromagnetic RF coils in high field MRI applications.

机译:在高场MRI应用中分析多杆横向电磁RF线圈的B1场轮廓和SAR值。

获取原文
获取原文并翻译 | 示例
       

摘要

In this work, the B1 field homogeneity and specific absorption rate (SAR) values were evaluated for three high-frequency (340 MHz) radio frequency coils designed for use in human magnetic resonance imaging at 8 tesla. Eight-, 16-, and 24-strut transverse electromagnetic (TEM) resonators were examined both experimentally and with the finite difference time domain numerical method. It was observed that increasing the number of TEM elements acted to lower the maximal achievable frequency of the coil and to increase the experimental complexities associated with tuning and matching. In addition, it is demonstrated from experiment and numerical analysis that the circularly polarized component of the B1 (B1+) field homogeneity in the head improved most from 8- to 16-strut coils. Numerical analysis revealed little difference in terms of SAR distribution between these coils; however, stronger tissue/coil coupling and consequently higher SAR peak values were obtained for the 8-strut case.
机译:在这项工作中,评估了设计用于8特斯拉人体磁共振成像的三个高频(340 MHz)射频线圈的B1场均匀性和比吸收率(SAR)值。分别通过实验和有限差分时域数值方法对八,十六和二十四杆横向电磁(TEM)谐振器进行了检查。观察到,增加TEM元素的数量可降低线圈的最大可实现频率,并增加与调谐和匹配相关的实验复杂性。此外,从实验和数值分析中可以看出,磁头中B1(B1 +)场均匀性的圆极化分量从8到16支线圈最大地改善了。数值分析表明,这些线圈之间的SAR分布几乎没有差异。但是,在8杆情况下,组织/线圈耦合更强,因此SAR峰值更高。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号