Dosimetric accuracy of a treatment planning system for actively scanned proton beams and small target volumes: Monte Carlo and experimental validation

机译:主动扫描质子束和小目标体积的治疗计划系统的剂量学准确性:蒙特卡洛和实验验证

获取原文
获取原文并翻译 | 示例
           

摘要

This study was performed to evaluate the accuracy of a commercial treatment planning system (TPS), in optimising proton pencil beam dose distributions for small targets of different sizes (5-30 mm side) located at increasing depths in water. The TPS analytical algorithm was benchmarked against experimental data and the FLUKA Monte Carlo (MC) code, previously validated for the selected beam-line. We tested the Siemens syngo (R) TPS plan optimisation module for water cubes fixing the configurable parameters at clinical standards, with homogeneous target coverage to a 2 Gy (RBE) dose prescription as unique goal. Plans were delivered and the dose at each volume centre was measured in water with a calibrated PTW Advanced Markus (R) chamber. An EBT3 (R) film was also positioned at the phantom entrance window for the acquisition of 2D dose maps. Discrepancies between TPS calculated and MC simulated values were mainly due to the different lateral spread modeling and resulted in being related to the field-to-spot size ratio. The accuracy of the TPS was proved to be clinically acceptable in all cases but very small and shallow volumes. In this contest, the use of MC to validate TPS results proved to be a reliable procedure for pre-treatment plan verification.
机译:进行这项研究以评估商业处理计划系统(TPS)在优化质子笔射束剂量分布方面的准确性,以优化位于水深处的不同大小(5-30 mm侧)的小目标。 TPS分析算法已针对实验数据和FLUKA蒙特卡洛(MC)代码进行了基准测试,之前已针对所选光束线进行了验证。我们测试了Siemens syngo(R)TPS计划优化模块的水立方,以固定临床标准上的可配置参数,并以2 Gy(RBE)剂量处方的均匀目标覆盖率作为唯一目标。交付计划并使用校准的PTW Advanced Markus(R)室在水中测量每个体积中心的剂量。还将EBT3(R)胶片放置在幻影入射窗处,以获取2D剂量图。 TPS计算值与MC模拟值之间的差异主要是由于侧向展布模型不同而导致的,并且与场对点尺寸比有关。在所有情况下,TPS的准确性在临床上都可以接受,但体积很小且很浅。在本次比赛中,使用MC验证TPS结果被证明是预处理方案验证的可靠程序。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号