首页> 外文期刊>Physics in medicine and biology. >An accurate calibration method of the multileaf collimator valid for conformal and intensity modulated radiation treatments
【24h】

An accurate calibration method of the multileaf collimator valid for conformal and intensity modulated radiation treatments

机译:一种多叶准直器的精确校准方法,适用于保形和强度调制的放射治疗

获取原文
获取原文并翻译 | 示例

摘要

Because for IMRT treatments the required accuracy on leaf positioning is high, conventional calibration methods may not be appropriate. The aim of this study was to develop the tools for an accurate MLC calibration valid for conventional and IMRT treatments and to investigate the stability of the MLC. A strip test consisting of nine adjacent segments 2 cm wide, separated by 1 mm and exposed on Kodak X-Omat V films at D_(max) depth, was used for detecting leaf-positioning errors. Dose profiles along the leaf-axis were taken for each leaf-pair. We measured the dose variation on each abutment to quantify the relative positioning error (RPE) and the absolute position of the abutment to quantify the absolute positioning error (APE). The accuracy of determining the APE and RPE was 0.15 and 0.04 mm, respectively. Using the RPE and the APE the MLC calibration parameters were calculated in order to obtain a flat profile on the abutment at the correct position. A conventionally calibrated Elekta MLC was re-calibrated using the strip test. The stability of the MLC and leaf-positioning reproducibility was investigated exposing films with 25 adjacent segments 1 cm wide during three months and measuring the standard deviation of the RPE values. A maximum shift over the three months of 0.27 mm was observed and the standard deviation of the RPE values was 0.11 mm.
机译:因为对于IMRT处理,所需的叶片定位精度很高,所以常规的校准方法可能不合适。这项研究的目的是开发一种适用于常规治疗和IMRT治疗的准确MLC校准工具,并研究MLC的稳定性。使用由9个2 cm宽,相距1 mm并在D_(max)深度处的Kodak X-Omat V胶片上曝光的相邻节段组成的试纸条来检测叶片位置误差。对于每个叶对,沿叶轴获取剂量分布。我们测量了每个基台的剂量变化,以量化相对定位误差(RPE)和基台的绝对位置,以量化绝对定位误差(APE)。确定APE和RPE的精度分别为0.15和0.04 mm。使用RPE和APE计算了MLC校准参数,以便在正确位置获得基台上的平坦轮廓。使用试纸条重新校准了常规校准的Elekta MLC。研究了MLC的稳定性和叶片定位的可重复性,在三个月内将25个相邻片段的宽度为1 cm的胶片曝光,并测量RPE值的标准偏差。观察到三个月的最大偏移为0.27 mm,RPE值的标准偏差为0.11 mm。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号