A method for determining the alignment accuracy of the treatment table axis at an isocentric irradiation facility.

机译:一种用于确定在等心辐照设施下治疗台轴对准精度的方法。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

At an isocentric irradiation facility, the rotation axis of the treatment table has to be accurately aligned in vertical orientation to the isocentre, which is usually marked by three perpendicular laser planes. In particular, high precision radiotherapy techniques, such as radiosurgery or intensity modulated radiotherapy, require a higher alignment accuracy of the table axis than routinely specified by the manufacturers. A simple and efficient method is presented to measure the direction and the size of the displacement of the table axis from the isocentre as marked by the lasers. In addition, the inclination of the table axis against the vertical direction can be determined. The measured displacement and inclination provide the required data to correct for possible misalignments of the treatment table axis and to maintain its alignment. Measurements were performed over a period of two years for a treatment table located at the German heavy ion therapy facility. The mean radial distance between the table axis and the isocentre was found to be 0.25 +/- 0.25 mm. The mean inclination of the table axis in the XZ- and YZ-planes was measured to be -0.03 +/- 0.02 degrees and -0.04 +/- 0.01 degrees, respectively. The measurements demonstrate the good alignment of the treatment table over the analysed time period. The described method can be applied to any isocentric irradiation facility, especially including isocentric linear accelerators used for radiosurgery or other high precision irradiation techniques.
机译:在等中心辐照设备上,治疗台的旋转轴必须在垂直方向上精确对准等中心点,等中心点通常由三个垂直的激光平面标记。特别地,诸如放射外科手术或强度调制放射疗法的高精度放射疗法技术需要比制造商常规规定的更高的工作台轴对准精度。提出了一种简单有效的方法来测量工作台轴相对于等中心线的位移方向和大小,如激光标记的那样。另外,可以确定工作台轴线相对于竖直方向的倾斜度。测得的位移和倾斜度可提供所需的数据,以校正治疗台轴可能的未对准并保持其对准。在德国重离子疗法设施的治疗台上进行了为期两年的测量。工作台轴与等中心点之间的平均径向距离为0.25 +/- 0.25毫米。测得的工作台轴在XZ和YZ平面上的平均倾斜度分别为-0.03 +/- 0.02度和-0.04 +/- 0.01度。测量结果表明在分析的时间段内治疗台的良好对准。所描述的方法可以应用于任何等中心辐射设施,特别是包括用于放射外科手术或其他高精度辐射技术的等中心线性加速器。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号