首页> 外文期刊>Physics in medicine and biology. >Design of ridge filters for spread-out Bragg peaks with Monte Carlo simulation in carbon ion therapy
【24h】

Design of ridge filters for spread-out Bragg peaks with Monte Carlo simulation in carbon ion therapy

机译:利用碳离子治疗中的蒙特卡洛模拟设计用于扩散布拉格峰的脊形滤波器

获取原文
获取原文并翻译 | 示例
           

摘要

Spread-out Bragg peaks made by ridge filters or wheel range modulators are used in charged particle therapy with passive methods to achieve uniform biological responses in irradiated tumors. Following the biological responses needed to design the ridge filters, which were developed at the National Institute of Radiological Sciences in Japan, new ridge filters were designed using recent developments in heavy-ion reactions and dosimetry. The Monte Carlo code of Geant4 was used to calculate the qualities of carbon ion beams in a water phantom. The results obtained from the simulation were corrected so that they agreed with the measurements of depth dose distributions. The calculations of biological responses to fragments other than carbon ions were assumed to be for helium ions. The measured dose distributions with the designed ridge filters were compared to the calculated distributions. A beam modifying system using this adaptable method was successively applied to carbon ion therapy at Gunma University.
机译:由脊形滤光片或轮距调节器产生的布拉格峰被用于带电粒子治疗中,采用被动方法,以在被照射的肿瘤中实现均匀的生物学反应。根据日本国家放射科学研究院开发的设计脊形过滤器所需的生物学响应,利用重离子反应和剂量学的最新进展设计了新的脊形过滤器。 Geant4的蒙特卡洛代码用于计算水模型中碳离子束的质量。从模拟获得的结果已得到校正,以使它们与深度剂量分布的测量结果一致。假定对除碳离子以外的碎片的生物响应的计算是针对氦离子。将使用设计的脊形过滤器测得的剂量分布与计算的分布进行比较。使用这种适应性方法的光束修改系统已在群马大学相继应用于碳离子治疗。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号