Feasibility of integrating a multi-camera optical tracking system in intra-operative electron radiation therapy scenarios

机译:在术中电子放射治疗方案中集成多摄像机光学跟踪系统的可行性

获取原文
获取原文并翻译 | 示例
           

摘要

Intra-operative electron radiation therapy (IOERT) combines surgery and ionizing radiation applied directly to an exposed unresected tumour mass or to a post-resection tumour bed. The radiation is collimated and conducted by a specific applicator docked to the linear accelerator. The dose distribution in tissues to be irradiated and in organs at risk can be planned through a pre-operative computed tomography (CT) study. However, surgical retraction of structures and resection of a tumour affecting normal tissues significantly modify the patient's geometry. Therefore, the treatment parameters (applicator dimension, pose (position and orientation), bevel angle, and beam energy) may require the original IOERT treatment plan to be modified depending on the actual surgical scenario. We propose the use of a multi-camera optical tracking system to reliably record the actual pose of the IOERT applicator in relation to the patient's anatomy in an environment prone to occlusion problems. This information can be integrated in the radio-surgical treatment planning system in order to generate a real-time accurate description of the IOERT scenario. We assessed the accuracy of the applicator pose by performing a phantom-based study that resembled three real clinical IOERT scenarios. The error obtained (2 mm) was below the acceptance threshold for external radiotherapy practice, thus encouraging future implementation of this approach in real clinical IOERT scenarios.
机译:术中电子放射疗法(IOERT)将手术和电离辐射直接应用于裸露的未切除肿瘤块或切除后的肿瘤床。辐射通过与线性加速器对接的专用施涂器进行准直和传导。可以通过术前计算机断层扫描(CT)研究来计划要照射的组织和有风险的器官中的剂量分布。然而,手术牵开结构和切除影响正常组织的肿瘤会显着改变患者的几何形状。因此,治疗参数(敷贴器尺寸,姿势(位置和方向),斜角和束能量)可能需要根据实际手术情况修改原始IOERT治疗计划。我们建议使用多相机光学跟踪系统,以在容易发生咬合问题的环境中可靠地记录IOERT施加器与患者解剖结构相关的实际姿势。可以将该信息集成到放射外科治疗计划系统中,以生成IOERT方案的实时准确描述。我们通过进行基于幻像的研究来评估涂药器姿势的准确性,该研究类似于三个实际的临床IOERT场景。所获得的误差(2毫米)低于外部放射疗法实践的接受阈值,因此鼓励了该方法在实际的临床IOERT方案中的进一步实施。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号