Performance analysis of a modified 3D dosimetry optical scanner based on the first generation optical CT scanner OCTOPUS is presented. The system consists of PRESAGE? dosimeters, the modified 3D scanner, and a new developed in-house user control panel written in Labview program which provides more flexibility to optimize mechanical control and data acquisition technique. The total scanning time has been significantly reduced from initial 8 h to ~2 h by using the modified scanner. The functional performance of the modified scanner has been evaluated in terms of the mechanical integrity uncertainty of the data acquisition process. Optical density distribution comparison between the modified scanner, OCTOPUS and the treatment plan system has been studied. It has been demonstrated that the agreement between the modified scanner and treatment plans is comparable with that between the OCTOPUS and treatment plans.
机译:提出了基于第一代光学CT扫描仪OCTOPUS的改进型3D剂量学光学扫描仪的性能分析。系统由PRESAGE组成?剂量计,改进的3D扫描仪以及用Labview程序编写的新开发的内部用户控制面板,可提供更大的灵活性来优化机械控制和数据采集技术。通过使用改进的扫描仪,总扫描时间从最初的8小时减少到了约2小时。已根据数据采集过程的机械完整性不确定性评估了改进型扫描仪的功能性能。研究了改进型扫描仪,OCTOPUS和治疗计划系统之间的光密度分布比较。事实证明,修改后的扫描仪与治疗计划之间的协议与OCTOPUS与治疗计划之间的协议具有可比性。