Diffuse reflectance spectrophotometry with visible light: comparison of four different methods in a tissue phantom

机译:可见光漫反射分光光度法:组织体模中四种不同方法的比较

获取原文
获取原文并翻译 | 示例
           

摘要

The purpose of the study was to compare algorithms of four methods (plus two modifications) for spectrophotometric haemoglobin saturation measurements. Comparison was made in tissue phantoms basically consisting of a phosphate buffer, Intralipid and blood, allowing samples to be taken for reference measurements. Three experimental series were made. In experiment A (eight phantoms) we used the Knoefel method and measured specific extinction coefficients with a reflection spectrophotometer. In experiment B (six phantoms) the fully oxygenated phantoms were gradually deoxygenated with baker's yeast, and simultaneous measurements were made with our spectrophotometer and with a reference oxymeter (ABL-605) in 3 min intervals. For each spectrophotometric measurement haemoglobin saturation was calculated with all algorithms and modifications, and compared with reference. In experiment C (11 phantoms) we evaluated the ability of a modification of the Knoefel method to measure haemoglobin concentration in absolute quantities using extinction coefficients from experiment A. Results. Experiment A: with the Knoefel method extinction coefficients (+/- SD) for oxyhaemoglobin at 553.04 and 573.75 nm were 1.117 (+/- 0.0396) ODmM(-1) and 1.680 (+/- 0.0815) ODmM(-1), respectively, and for deoxyhaemoglobin 1.205 (+/- 0.0514) ODmM(-1) and 0.953 (+/- 0.0487) ODmM-1, respectively. Experiment B: high correlation with the reference was found in all methods (r = 0.94-0.97). However, agreement varied from evidently wrong in method 3 and the original method 4 (e.g. saturation above 160%) to high agreement in method 2 as well as the modifications of methods 1 and 4, where oxygen dissociation curves were close to the reference method. Experiment C: with the modified Knoefel method the mean haemoglobin concentration difference from reference was 8.3% and the correlation was high (r = 0.91). We conclude that method 2 and the modifications of 1 and 4 were superior to the others, but depended on known values in the same or similar phantoms. The original method I was independent of results from the tissue phantoms, but agreement was slightly poorer. Method 3 and the original method 4 could not be recommended. The ability of the modified method 1 to measure haemoglobin concentration is promising, but needs further development.
机译:本研究的目的是比较分光光度法血红蛋白饱和度测量的四种方法(加上两种修改)的算法。在基本由磷酸盐缓冲液,脂质内和血液组成的组织体模中进行了比较,从而可以将样品用于参考测量。进行了三个实验系列。在实验A(八个模型)中,我们使用Knoefel方法,并使用反射分光光度计测量了特定的消光系数。在实验B(六个模型)中,完全氧化的模型用面包酵母逐渐脱氧,并用我们的分光光度计和参考血氧仪(ABL-605)每隔3分钟进行一次同时测量。对于每个分光光度法测量,均使用所有算法和修改方法计算血红蛋白饱和度,并与参考值进行比较。在实验C(11个幻像)中,我们使用实验A的消光系数评估了Knoefel方法的改进,以绝对量测量血红蛋白浓度的能力。结果。实验A:用Knoefel方法在553.04和573.75 nm处血红蛋白的消光系数(+/- SD)分别为1.117(+/- 0.0396)ODmM(-1)和1.680(+/- 0.0815)ODmM(-1) ,而对于脱氧血红蛋白则分别为1.205(+/- 0.0514)ODmM(-1)和0.953(+/- 0.0487)ODmM-1。实验B:在所有方法中均发现与参考物高度相关(r = 0.94-0.97)。但是,从方法3和原始方法4的明显错误(例如饱和度超过160%)到方法2的高度一致性以及方法1和4的修改(其中氧解离曲线接近于参考方法),一致性有所不同。实验C:采用改良的Knoefel方法,与参考相比平均血红蛋白浓度差异为8.3%,相关性很高(r = 0.91)。我们得出的结论是,方法2以及方法1和4的修改优于其他方法,但是取决于相同或相似模型中的已知值。我最初的方法与组织体模的结果无关,但是一致性稍差。不推荐使用方法3和原始方法4。改进方法1测量血红蛋白浓度的能力是有前途的,但还需要进一步发展。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号