首页> 外文期刊>Physics in medicine and biology. >Irradiating cell samples in an x-ray radiation cabinet—the effect of tube filtration
【24h】

Irradiating cell samples in an x-ray radiation cabinet—the effect of tube filtration

机译:在X射线辐射柜中辐射细胞样本-管过滤的效果

获取原文
获取原文并翻译 | 示例
           

摘要

The cell irradiation cabinet described is used for creating DNA damage in cell samples in order to study tumourigenesis. The medical research laboratory involved was using the manufacturer's quoted dose rate (32.2 mGy s~(-1)) to determine the required exposure time to impart a dose of 10 Gy. The x-ray output was investigated when the exposure failed to produce cell cycle arrest. The x-ray tube was fitted with only a 0.76 mm beryllium filter, and the spectrum therefore contained a high proportion of low energy photons which were being removed by the polystyrene sample flask as demonstrated by dose measurements in air and through the sample flask. Incorporation of a 0.5 mm aluminium filter removed most of these low energy photons but greatly reduced the dose rate to 3.8 mGy s~(-1). The manufacturer's quoted dose rates from a lightly filtered tube are misleading: the contribution of the very low energy component of the spectrum to the dose is not relevant since it will be removed by a sample flask.
机译:所描述的细胞辐照柜用于在细胞样品中造成DNA损伤,以研究肿瘤发生。所涉及的医学研究实验室正在使用制造商的报价剂量率(32.2 mGy s〜(-1))确定给予10 Gy剂量所需的暴露时间。当暴露未能产生细胞周期停滞时,研究X射线输出。 X射线管仅装有一个0.76毫米的铍滤光片,因此该光谱包含高比例的低能光子,这些低能光子已被聚苯乙烯样品瓶除去,如通过空气中和通过样品瓶中的剂量测量所证明的。引入0.5 mm的铝滤光片可去除大部分这些低能光子,但剂量率却大大降低至3.8 mGy s〜(-1)。制造商从轻过滤管中引用的剂量率具有误导性:光谱中非常低的能量成分对剂量的影响不大,因为它将被样品瓶除去。