首页> 外文期刊>Physics in medicine and biology. >Reply to comment on 'Proton beam monitor chamber calibration'
【24h】

Reply to comment on 'Proton beam monitor chamber calibration'

机译:回复对“质子束监控室校准”的评论

获取原文
获取原文并翻译 | 示例
           

摘要

This reply shows that the discrepancy of about 3% between Faraday cup dosimetry and reference dosimetry using a cylindrical ionization chamber found in Goma (2014 Phys. Med. Biol. 59 4961-71) seems to be due to an overestimation of the beam quality correction factors tabulated in IAEA TRS-398 for the cylindrical chamber used, rather than to 'unresolved problems with Faraday cup dosimetry', as suggested by Palmans and Vatnitsky (2016 Phys. Med. Biol. 61 6585-93). Furthermore, this work shows that a good agreement between reference dosimetry and Faraday cup dosimetry is possible, provided accurate beam quality correction factors for proton beams are used. The review on W-air values presented by Palmans and Vatnitsky is believed to be inaccurate, as it is based on the imprecise assumption of ionization chamber perturbation correction factors in proton beams being equal to unity.
机译:该答复表明,在戈马发现的法拉第杯剂量法和使用圆柱形电离室的参考剂量法之间存在约3%的差异(2014 Phys。Med。Biol。59 4961-71)似乎是由于对射束质量校正的过高估计如Palmans和Vatnitsky(2016 Phys.Med.Biol.61 6585-93)所建议,在IAEA TRS-398中列出了所用圆柱形腔室的一些因素,而不是``未解决的法拉第杯剂量法问题''。此外,这项工作表明,只要使用质子束的精确束质量校正因子,参考剂量法和法拉第杯剂量法之间的良好一致性是可能的。人们认为,Palmans和Vatnitsky对W-air值的审查是不准确的,因为它是基于质子束中电离室扰动校正因子等于1的不精确假设。