The accuracy of image registration for the brain and the nasopharynx using external anatomical landmarks.

机译:使用外部解剖界标对大脑和鼻咽进行图像配准的准确性。

获取原文
获取原文并翻译 | 示例

摘要

We investigated the accuracy of 3D image registration using markers that are repeatedly applied to external anatomical landmarks on the head. The purpose of this study is to establish a lower limit of the errors that would occur in, for instance, MRI-SPECT matching, which in some situations can only be achieved using external landmarks. Marker matching was compared with (single-modality) volume matching for 20 MRI scans. The results were compared with a published expression for the target registration error (TRE) which gives the 3D distribution of the mismatch between both scans. It was found that the main error source is reapplying the external markers on the anatomical landmarks. The published expression describes the relative distribution of the TRE in space well, but tends to underestimate the actual registration error. This deviation is due to anisotropy in the error distribution of the marker position (errors in the direction perpendicular to the skin surface are in general much smaller than errors in other directions). A simulation of marker matching with anisotropy in the errors confirmed this finding. With four reapplied markers, the TRE is 6 mm or smaller in most regions of the head.
机译:我们研究了使用标记的3D图像配准的准确性,这些标记被反复应用于头部的外部解剖界标。这项研究的目的是确定例如MRI-SPECT匹配中将发生的错误的下限,在某些情况下只能使用外部界标来实现。将标记匹配与(单模式)体积匹配进行20次MRI扫描。将结果与目标配准误差(TRE)的已发布表达式进行比较,该表达式给出了两次扫描之间不匹配的3D分布。发现主要错误源是在解剖界标上重新应用外部标记。公开发表的表述很好地描述了TRE在空间中的相对分布,但往往低估了实际配准误差。该偏差归因于标记器位置的误差分布中的各向异性(与皮肤表面垂直的方向上的误差通常比其他方向上的误差小得多)。在误差中具有各向异性的标记匹配的模拟证实了这一发现。有了四个重新应用的标记,TRE在头部的大部分区域为6 mm或更小。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号