Toward acquiring comprehensive radiosurgery field commissioning data using the PRESAGE/optical-CT 3D dosimetry system.

机译:希望使用PRESAGE / optical-CT 3D剂量系统获取全面的放射外科现场调试数据。

获取原文
获取原文并翻译 | 示例

摘要

Achieving accurate small field dosimetry is challenging. This study investigates the utility of a radiochromic plastic PRESAGE read with optical-CT for the acquisition of radiosurgery field commissioning data from a Novalis Tx system with a high-definition multileaf collimator (HDMLC). Total scatter factors (Sc, p), beam profiles, and penumbrae were measured for five different radiosurgery fields (5, 10, 20, 30 and 40 mm) using a commercially available optical-CT scanner (OCTOPUS, MGS Research). The percent depth dose (PDD), beam profile and penumbra of the 10 mm field were also measured using a higher resolution in-house prototype CCD-based scanner. Gafchromic EBT film was used for independent verification. Measurements of Sc, p made with PRESAGE and film agreed with mini-ion chamber commissioning data to within 4% for every field (range 0.2-3.6% for PRESAGE, and 1.6-3.6% for EBT). PDD, beam profile and penumbra measurements made with the two PRESAGE/optical-CT systems and film showed good agreement with the high-resolution diode commissioning measurements with a competitive resolution (0.5 mm pixels). The in-house prototype optical-CT scanner allowed much finer resolution compared with previous applications of PRESAGE. The advantages of the PRESAGE system for small field dosimetry include 3D measurements, negligible volume averaging, directional insensitivity, an absence of beam perturbations, energy and dose rate independence.
机译:实现精确的小剂量现场测量是具有挑战性的。这项研究调查了使用光学CT读取的放射性变色塑料PRESAGE在具有高清晰度多叶准直仪(HDMLC)的Novalis Tx系统中获取放射外科现场调试数据的实用性。使用市售的光学CT扫描仪(OCTOPUS,MGS Research)测量了五个不同的放射外科手术区域(5、10、20、30和40 mm)的总散射因子(Sc,p),光束轮廓和半影。还使用高分辨率的内部原型基于CCD的扫描仪测量了10 mm视野的百分比深度剂量(PDD),光束轮廓和半影。全色EBT膜用于独立验证。用PRESAGE和胶片进行的Sc,p的测量值与小型离子室调试数据一致,每个场均在4%以内(PRESAGE范围为0.2-3.6%,EBT范围为1.6-3.6%)。使用两个PRESAGE / optical-CT系统和胶片进行的PDD,光束轮廓和半影测量与高分辨率二极管调试测量(分辨率为0.5 mm像素)具有很好的一致性。与PRESAGE以前的应用相比,内部原型光学CT扫描仪具有更高的分辨率。用于小场剂量测定的PRESAGE系统的优点包括3D测量,可忽略的体积平均,方向不敏感,不存在束扰,能量和剂量率独立性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号