PURPOSE: For a novel needle-free injection (NFI) system, the relationship between frequency of wet or incomplete injections and device-related factors and subject physiological variables was examined. MATERIALS AND METHODS: A total of 26 device configurations of a single-use pre-filled NFI system (Intraject) were used to deliver a total of 3,211 subcutaneous injections into the abdomen of 302 healthy volunteers. Two validated methods were used to determine completeness of each injection (defined as >or=90% dose delivery). Skin-fold thickness, body mass index (BMI), Fitzpatrick skin type, sex, age, and injection site were noted for each volunteer. RESULTS: The proportion of complete injections ranged from 59-98% among the various combinations of device configurations. Two device parameters and two subject-related variables showed strong association with injection performance; Device gas mass (chamber pressure) and orifice size demonstrated statistically significant, independent effects, with increasing gas mass and larger orifice size associated with improved injection performance. BMI and site of injection on the abdomen also demonstrated statistically significant effects with increasing BMI and lateral rather than medial injection sites associated with better injections. CONCLUSION: Both device-related factors and subject variables interact to mediate in vivo performance of a needle-free injector.
机译:目的:对于新型无针注射(NFI)系统,检查了湿式或不完全注射的频率与设备相关因素与受试者生理变量之间的关系。材料和方法:一次性使用预填充NFI系统(Intraject)的总共26种设备配置用于向302名健康志愿者的腹部中总共进行3,211例皮下注射。使用两种经过验证的方法确定每次注射的完整性(定义为>或= 90%的剂量输送)。记录每位志愿者的皮肤厚度,体重指数(BMI),费兹帕特里克皮肤类型,性别,年龄和注射部位。结果:在设备配置的各种组合中,完全注射的比例在59%至98%之间。两个设备参数和两个与受试者相关的变量与注射性能密切相关。装置气体质量(腔室压力)和孔尺寸在统计学上具有显着的独立影响,随着气体质量的增加和更大的孔尺寸与改进的注射性能相关。 BMI和腹部注射部位还显示出统计学上显着的效果,即BMI和外侧注射部位的增加,而不是与更好注射相关的内侧注射部位的增加。结论:与设备有关的因素和受试者变量相互作用,以介导无针注射器的体内性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号