Differences in displayed pump flow compared to measured flow under varying conditions during simulated cardiopulmonary bypass.

机译:在模拟心肺旁路过程中,在变化的条件下,显示的泵流量与测量流量的差异。

获取原文
获取原文并翻译 | 示例
           

摘要

Errors in blood flow delivery due to shunting have been reported to reduce flow by, potentially, up to 40-83% during cardiopulmonary bypass. The standard roller-pump measures revolutions per minute and a calibration factor for different tubing sizes calculates and displays flow accordingly. We compared displayed roller-pump flow with ultrasonically measured flow to ascertain if measured flow correlated with the heart-lung pump flow reading. Comparison of flows was measured under varying conditions of pump run duration, temperature, viscosity, varying arterial/venous loops, occlusiveness, outlet pressure, use of silicone or polyvinyl chloride (PVC) in the roller race, different tubing diameters, and use of a venous vacuum-drainage device.
机译:据报道,由于分流造成的血流输送错误会在体外循环期间将血流减少多达40-83%。标准的滚筒泵测量每分钟的转数,并针对不同尺寸的管子计算校准系数,并相应地显示流量。我们将显示的滚子泵流量与超声测量流量进行比较,以确定测量流量是否与心肺泵流量读数相关。在不同的泵运行时间,温度,粘度,变化的动脉/静脉回路,闭塞性,出口压力,在滚道中使用有机硅或聚氯乙烯(PVC),不同管径和使用静脉真空引流装置。