Significant reduction of air microbubbles with the dynamic bubble trap during cardiopulmonary bypass.

机译:在体外循环过程中通过动态气泡收集器大大减少了空气微气泡。

获取原文
获取原文并翻译 | 示例
       

摘要

Air microbubbles mostly occur unnoticed during cardiopulmonary bypass and are predominantly responsible for serious postoperative psychoneurological dysfunction. A dynamic bubble trap (DBT), which removes air microbubbles from the arterial blood, was tested in a clinical study. The aim was to evaluate the efficiency of microbubble removal under clinical conditions. As blood passes through the DBT, which is placed in the arterial line between the arterial filter and arterial cannula, it is converted into a rotating stream. The bubbles are directed to the centre of the blood flow and are collected in the distal end of the DBT, from where they are returned to the cardiotomy reservoir. Doppler ultrasonography was used to detect the microbubbles before and after the DBT, and also the number of high-intensity transient signals (HITS) in the right and left middle cerebral artery during extracorporeal circulation. A significant reduction of microbubbles in the arterial line (3,990 before DBT, 537 after, p < 0.001) and HITS in the brain (51 in the DBT group, 77 in the placebo group, p = 0.04) was measured.
机译:空气微气泡多数在心肺转流过程中未引起注意,并且是导致严重的术后心理神经功能障碍的主要原因。在临床研究中测试了一种动态气泡收集器(DBT),它可以去除动脉血中的空气微气泡。目的是评估临床条件下去除微泡的效率。当血液通过DBT时,该DBT置于动脉过滤器和动脉套管之间的动脉管线中,它被转化为旋转流。气泡被引导到血流的中心,并被收集在DBT的远端,然后从那里返回到心脏切开术储液器。使用多普勒超声检查来检测DBT前后的微气泡,以及体外循环期间左右大脑中动脉的高强度瞬时信号(HITS)的数量。测得动脉线微泡(DBT前为3,990,DBT后为537,p <0.001)和HITS(DBT组为51,安慰剂组为77,p = 0.04)显着减少。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号