首页> 外文期刊>Perfusion >Quantitative haemodynamic evaluation of aortic cannulas.
【24h】

Quantitative haemodynamic evaluation of aortic cannulas.

机译:主动脉插管的定量血流动力学评估。

获取原文
获取原文并翻译 | 示例
       

摘要

We have designed a new type of aortic cannula for extracorporeal circulation which we hypothesize has superior haemodynamic characteristics compared with existing types. The aim of the study was to perform a quantitative evaluation of the haemodynamic performance of commercially available aortic cannulas against our new design. Six existing cannulas and two self-designed, modified DLP cannulas were investigated. Pressure drop, peak velocity and wall thickness were obtained from in vitro experiments. Pressure drop and turbulent stresses were obtained by in vivo experiments. At 5 L/min the pressure drops of the new design were 18-25 mmHg in vitro and 20 mmHg in vivo compared with 25-92 mmHg in vitro and 28-90 mmHg in vivo for the six commercially available cannulas. Turbulent stresses of the new design varied between 25 and 50 N/m2 in vivo compared with 125-500 N/m2. Both in vivo and in vitro results showed a clear haemodynamic superiority of the new cannula-tip design compared with all existing types.
机译:我们设计了一种新型的用于体外循环的主动脉插管,我们认为它比现有的具有更高的血液动力学特性。这项研究的目的是针对我们的新设计对商用主动脉插管的血液动力学性能进行定量评估。研究了六个现有插管和两个自行设计的改良DLP插管。从体外实验获得压降,峰值速度和壁厚。通过体内实验获得压降和湍流应力。新设计的压力为5 L / min时,体外的18-25 mmHg和体内的20 mmHg压降,而六个商用插管的体外25-92 mmHg和28-90 mmHg的压降。相比于125-500 N / m2,新设计的体内湍流应力在25至50 N / m2之间变化。体内和体外结果均显示,与所有现有类型相比,新的针尖设计具有明显的血液动力学优势。