Developing low-dose C-arm CT imaging for temporomandibular joint (TMJ) disorder in interventional radiology.

机译:在介入放射学中开发用于颞下颌关节(TMJ)疾病的小剂量C型臂CT成像。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: Manufacturers have provided C-arm CT imaging technologies for applications in interventional radiology in recent years. However, clinical imaging protocols and radiation doses have not been well studied or reported. OBJECTIVE: The purpose of this study is to develop low-dose settings for clinically acceptable CT imaging of temporomandibular joint in interventional radiology suites, using a C-arm imaging angiography system. MATERIALS AND METHODS: CT scans were performed with a flat-panel digital C-arm angiographic system on a 5-year-old anthropomorphic phantom. The CTDI was determined for various rotation times, dose settings and Cu filter selections. The CTDI values were compared with those of conventional low-dose CT for the same phantom. The effectiveness of using Cu filters to reduce dose was also investigated. Images were reviewed by a senior radiologist for clinical acceptance. RESULTS: The manufacturer's default setting gave an equivalent CTDI of 4.8 mGy. Optimizing the dose settings and adding copper filtration reduced the radiation dose by 94%. This represents a 50% reduction from conventional CT. CONCLUSION: Use of Cu filters and low-dose settings significantly reduced radiation dose from that of standard settings. This phantom study process successfully guided the clinical implementation of low-dose studies for all ages at our institution.
机译:背景:近年来,制造商已提供C臂CT成像技术,用于介入放射学。但是,临床成像方案和辐射剂量尚未得到很好的研究或报道。目的:本研究的目的是利用C型臂成像血管造影系统,为介入放射学室开发颞下颌关节CT成像的低剂量设置。材料与方法:CT扫描是在5岁的拟人化体模上使用平板数字C型臂血管造影系统进行的。确定了各种旋转时间,剂量设置和铜过滤器选择的CTDI。对于相同的体模,将CTDI值与常规低剂量CT的CTDI值进行比较。还研究了使用铜过滤器降低剂量的有效性。图像由一位高级放射科医生审查,以用于临床。结果:制造商的默认设置给出了4.8 mGy的等效CTDI。优化剂量设置并添加铜过滤可将辐射剂量降低94%。与传统的CT相比减少了50%。结论:使用铜过滤器和低剂量设置可显着降低标准剂量的辐射剂量。这种幻象研究过程成功指导了我们机构所有年龄段小剂量研究的临床实施。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号