Interrupter resistance (R(int)) is a useful measure of airway caliber in young children, but has not been well characterized in infants-in whom there are concerns about the accurate measurement of driving pressure. This study aimed to assess the feasibility and repeatability of measuring R(int) in unsedated newborn infants, and to explore alternative algorithms for calculating driving pressure. R(int) measurement was attempted in 28 healthy term newborn infants during natural sleep using the MicroRint device. Paired R(int) measurements were achieved in 24 infants, but after screening of waveforms only 15 infants had at least 5 technically acceptable waveforms on both measurements. R(int) values obtained were comparable with reported values for airflow resistance in newborns using other methods. However, the repeatability coefficient (CR) was much higher than reported values in preschool children using standard back-extrapolation algorithms, with CR 2.47 KPa L(-1) sec (unscreened) and 2.93 KPa L(-1) sec (screened). Other algorithms gave only marginally better repeatability, with all CR values over 50% of the mean R(int) value. Using current commercially available equipment, R(int) is too poorly repeatable to be a reliable measurement of airflow resistance in newborn infants. Lower deadspace equipment is needed, but anatomical and physiological factors in the infant are also important.
机译:灭弧室阻力(R(int))是衡量幼儿气道口径的一种有用方法,但对于婴儿(对精确测量驱动压力存在担忧),其特征尚不明确。这项研究旨在评估在未镇静的新生儿中测量R(int)的可行性和可重复性,并探索计算驾驶压力的替代算法。使用MicroRint设备在自然睡眠期间尝试对28名健康足月新生儿进行R(int)测量。配对R(int)测量在24例婴儿中完成,但在筛选波形后,只有15例婴儿在两次测量中均具有至少5个技术上可接受的波形。使用其他方法获得的R(int)值与报道的新生儿气流阻力值相当。但是,使用标准的反向外推算法,学龄前儿童的可重复性系数(CR)远远高于报告值,CR为2.47 KPa L(-1)sec(未筛选)和2.93 KPa L(-1)sec(筛选)。其他算法仅在一定程度上提高了可重复性,所有CR值均超过平均R(int)值的50%。使用当前的市售设备,R(int)的可重复性太差,无法可靠地测量新生儿的气流阻力。需要较低的死角设备,但婴儿的解剖和生理因素也很重要。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号