A pilot validation study for the NOX T3(TM) portable monitor for the detection of OSA.

机译:对用于检测OSA的NOX T3™便携式监控器的试点验证研究。

获取原文
获取原文并翻译 | 示例
           

摘要

In this small, clinic-based sample, the T3 demonstrated very good measurement agreement compared to PSG and a high degree of sensitivity for detecting even mild OSA. False positives appeared to be due to respiratory effort-related arousals (RERAs) being autoscored as obstructive apneas and may be due to inherent discrepancy in flow measurement sensitivity between PSG and portable monitors.
机译:在这个基于临床的小样本中,与PSG相比,T3表现出非常好的测量一致性,并且即使检测到轻微的OSA也具有很高的灵敏度。误报似乎是由于呼吸努力相关唤醒(RERA)被自动识别为阻塞性呼吸暂停,并且可能是由于PSG与便携式监护仪之间流量测量灵敏度的固有差异。