Optimized pulse durations minimize the effect of polarity reversal on defibrillation efficacy with biphasic shocks.

机译:优化的脉冲持续时间可最大程度地减小极性反转对双相电击除颤功效的影响。

获取原文
获取原文并翻译 | 示例
       

摘要

There are conflicting results on the effect of polarity change on the defibrillation efficacy of biphasic shocks possibly caused by different shock durations. The goal of the present study was to investigate the influence of polarity reversal on defibrillation efficacy for different biphasic shock durations in a porcine animal model. In eight anesthesized pigs using a transvenous/submuscular lead system DFTs for 4 phase 1 durations were determined: 8.1 ms, 6 ms, 3.8 ms and 1.7 ms. The phase 1/phase 2 ratio was constant at 60%/40%. For cathodal shocks, the defibrillation coil in the right ventricular apex was the cathode during phase 1 and for anodal shocks it was the anode. For both polarities, the strength-duration curve revealed a DFT minimum at 3.8 ms (cathodal shocks: 21.3 +/- 6.4 J, P < 0.001; anodal shocks: 21.9 +/- 8 J, P = 0.05). For anodal shocks and phase 1 durations of 1.7, 3.8, and 6 ms there was no significant difference of the stored energy at the DFT compared to cathodal shocks. In contrast, significantly lower DFTs were observed for anodal shocks with a phase 1 duration of 8.1 ms (28.8 +/- 6.4 J compared to 33.1 +/- 5.9 J for cathodal shocks, P = 0.006). The effect of lower defibrillation energy requirements with polarity reversal depends on the total biphasic shock duration; for the pulse duration with the lowest DFT, polarity reversal does not increase defibrillation efficacy of biphasic shocks.
机译:极性变化对双相电击除颤效果的影响存在矛盾的结果,这可能是由于电击持续时间不同而引起的。本研究的目的是在猪动物模型中研究极性反转对除颤效果的影响。在八只使用经静脉/肌下铅系统的麻醉猪中,确定了4个1期持续时间的DFT:8.1 ms,6 ms,3.8 ms和1.7 ms。相1 /相2的比率恒定在60%/ 40%。对于阴极电击,在第1阶段中,右心室心尖的除颤线圈为阴极,而对于阳极电击,则为阳极。对于两种极性,强度-持续时间曲线均显示在3.8毫秒时的DFT最小值(阴极冲击:21.3 +/- 6.4 J,P <0.001;阳极冲击:21.9 +/- 8 J,P = 0.05)。对于阳极电击和阶段1持续时间1.7、3.8和6毫秒,与阴极电击相比,DFT处的存储能量没有显着差异。相反,对于阳极电击,第1阶段持续时间为8.1毫秒(28.8 +/- 6.4 J,而阴极电击为33.1 +/- 5.9 J,P = 0.006),观察到的DFT明显较低。极性反转时降低除颤能量要求的效果取决于总的双相电击持续时间。对于DFT最低的脉冲持续时间,极性反转不会增加双相电击的除颤功效。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号