Strength duration curve for left ventricular epicardial stimulation in patients undergoing cardiac resynchronization therapy.

机译:接受心脏再同步治疗的患者左心室心外膜刺激的强度持续时间曲线。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

INTRODUCTION: The strength duration curve has been studied for right ventricular endocardial stimulation. There are differences between left ventricular epicardial and right ventricular endocardial stimulation due to different electrophysiologic properties and different electrode-tissue interface. The strength duration curve for epicardial left ventricular stimulation has not been studied so far. METHODS: One hundred and three patients were studied. The strength duration curves were determined for left ventricular epicardial and right ventricular endocardial stimulation. The studied points were chronaxie, rheobase, and voltage threshold at 0.5 ms. Left ventricular leads Guidant 4512, 4513, 4537, 4538 (unipolar, area 3.5 mm(2); Guidant Corp., St. Paul, MN, USA), Medtronic 4193 (unipolar, area 5.8 mm(2); Medtronic Inc., Minneapolis, MN, USA), Guidant 4518, 4542, 4543 (bipolar, area 4 mm(2)), St. Jude Medical (bipolar, area 4.8 mm(2); St. Jude Medical, St. Paul, MN, USA), and Medtronic 4194 (bipolar, area 5.8 mm(2)) were studied. RESULTS: The Guidant unipolar leads with a distal electrode area of 3.5 mm(2) had a lower chronaxie than the other studied leads. The left ventricular epicardial and right ventricular endocardial chronaxie for 15 patients with Medtronic left ventricular leads 4194 or 4193 (5.8 mm(2)) and right ventricular leads 6947 (5.7 mm(2)) were 0.52 +/- 0.36 ms and 0.62 +/- 0.46 ms (P > 0.05). CONCLUSION: The left ventricular epicardial chronaxie depends on the lead. The left ventricular epicardial chronaxie is similar to the right ventricular endocardial chronaxie for leads with similar electrode stimulation area.
机译:简介:研究了右心室心内膜刺激的强度持续时间曲线。由于不同的电生理特性和不同的电极-组织界面,左心室心外膜和右心室心内膜刺激之间存在差异。迄今为止尚未研究心外膜左心室刺激的强度持续时间曲线。方法:对103例患者进行了研究。确定左心室心外膜和右心室心内膜刺激的强度持续时间曲线。研究的点是时差,流变基础和0.5 ms的电压阈值。左心室导线Guidant 4512、4513、4537、4538(单极,面积3.5 mm(2); Guidant Corp.,圣保罗,美国明尼苏达州),Medtronic 4193(单极,面积5.8 mm(2); Medtronic Inc.美国明尼苏达州明尼阿波利斯),Guidant 4518、4542、4543(双极,面积4 mm(2)),圣裘德医疗(双极,面积4.8 mm(2);圣裘德医学,美国明尼苏达州圣保罗)和Medtronic 4194(双极型,面积5.8 mm(2))进行了研究。结果:远端电极面积为3.5 mm(2)的Guidant单极导线比其他研究的导线具有更低的时轴。 Medtronic左心室导线4194或4193(5.8 mm(2))和右心室导线6947(5.7 mm(2))的15例患者的左室心外膜和右室心内膜时轴分别为0.52 +/- 0.36 ms和0.62 + / -0.46毫秒(P> 0.05)。结论:左心室心外膜时轴取决于导线。对于具有相似电极刺激面积的导线,左心室心外膜时轴类似于右心室心内膜时轴。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号