Isorhythmic dissociation is an uncommon but clinically important finding in a patient implanted with a dual chamber cardioverter defibrillator (ICD) capable of SmartSensing. The surface ECG mimicked a malfunction of the pacemaker that was related to the concomitant hemodynamic deterioration. Device interrogation revealed dissociation of the intrinsic junctional and the pacemaker rhythm, which appeared at similar rates. Careful analysis of intracardiac electrograms and marker channel annotations helped to unravel the limitations for physiologic pacing.
机译:在植入了具有SmartSensing功能的双室心脏复律除颤器(ICD)的患者中,等速心律分离是罕见的,但在临床上很重要。表面心电图模拟了起搏器的故障,该故障与伴随的血液动力学恶化有关。装置询问显示出固有的连接节律和起搏器节律分离,这以相似的速率出现。心内电描记图和标记通道注释的仔细分析有助于揭示生理起搏的局限性。